Induction Seal Liner Peel Strength Tester
| Product code | CMI-IND-0197 |
| Categories | Plastic Caps and Closure Testing Instruments |
| Standards | Manufactured to IS / BS / EN / ASTM / AASHTO as applicable |
| Supply | India and export — 90+ countries |
- Description
Peel tester with bottle holder to measure the force needed to remove an induction sealed foil liner from a container neck. Used by pharma and food bottlers.
Specification
A bottle neck with the foil seal still attached is clamped and the foil is pulled at 90 degrees while a load cell records force. Too low a value risks leaking and too high makes opening difficult. The test helps set the induction sealer power.
Features
- Load cell 100 N
- 90 degree peel fixture
- Neck clamps for 28 - 63 mm
- Speed 50 - 300 mm/min
- Peak and average force
Technical Specifications
| Capacity | 100 N |
| Resolution | 0.01 N |
| Peel Speed | 50 - 300 mm/min |
| Neck Size | 28 - 63 mm |
| Peel Angle | 90 deg |
| Standards | ASTM F904 (principle), ASTM D903 |
| Power Supply | 230 V AC, 50 Hz |
Standard Accessories
- Neck clamps
- Foil grips
Induction Seal Liner Peel Strength Tester suppliers in India